SW-60&SW-61  

Syringe Heaters

Warner

 

SW-60

SW-60 Syringe Heater

Solution reservoir heating is an important technique used to eliminate outgassing of solutions prior to their entry to a heated perfusion chamber.

  • Compatible with both the PI-1 Power Interface and SW-707 Power Controller
  • Options for both single master syringe temperature and multiple independent syringe temperatures
  • Designed for situations where a large heated water bath is inconvenient (such as a shared facility)
  • Designed for use with BD plastic syringes

 



Solution reservoir heating is an important technique used to eliminate outgassing of solutions prior to their entry to a heated perfusion chamber. Since the gas load of a solution is dependent on partial pressure and temperature, preheating the solution at atmospheric pressure before delivery to the final heater will minimize the occurrence of bubbles in the chamber's bath, even if the solution is allowed to cool en route.

Solution reservoir heaters from Warner are designed for applications where the use of a large heated water bath is inconvenient, such as a shared facility. The ability to independently control each separate heater block allows the researcher to control the initial temperature of each solution without influencing other nearby solutions.

The SW-60 accommodates a single 60 cc syringe and its housing is made of anodized aluminum which is both corrosion resistant and serves as an excellent conductor of heat. These heaters may be mounted on any 3/8 inch diameter lab rod or used with the PS-560 stand.

A TA-29 thermistor is supplied for monitoring the temperature within the syringe. A 60 cc syringe is not supplied with this product. Syringes may be ordered separately. The syringe warmer will fit a BD size syringe. Additional useful accessories are also available.

 

 

SW-61

SW-61 Syringe Heater

Solution reservoir heating is an important technique used to eliminate outgassing of perfusate solutions.

  • Compatible with TC-324 and TC-344 Heater Controllers
  • Quiet operation
  • Compact design accommodates a single 60 cc syringe
  • Designed for situations where a large heated water bath is inconvenient (such as a shared facility) 

 



Solution reservoir heating is an important technique used to eliminate outgassing of perfustae solutions prior to their entry to a heated perfusion chamber. Since the gas load of a solution has dependence on pressure and temperature, preheating a solution at the final pressure before delivery to the chamber will minimize the occurrence of bubbles in the chamber's bath, even if the solution is allowed to cool en route.

Solution reservoir heaters from Warner Instruments are designed for applications where the use of a large heated water bath is inconvenient, such as a shared facility. Designed to accommodate one 60 cc syringe, the SW-61 Heater housing is made of anodized aluminum which is both corrosion resistant and serves as an excellent conductor. Solutions will reach set temperature approximately 15 minutes after application of power.

These heaters may be mounted on any 3/8" or 1/2" diameter lab rod and each unit is supplied with a cable assembly for connection to Warner's TC-324 or TC-344 heater controllers (purchased separately). A TA-29 thermistor is supplied for monitoring the temperature within the syringe. A 60 cc syringe is not supplied with this product. Syringes may be ordered separately. The syringe warmer will fit a BD size syringe. Additional useful accessories are also available. 

 

                                  
                                Specifications

SW-60 SW-61

Syringe Size

60 cc

60 cc

Heater Resistance

6.67Ω

 10Ω

Voltage Requirement

Input Voltage 12 V nominal, 16 V max

Variable to 12 V

Max. Temperature Range

25 - 50°C

25 - 50 °C

Temperature Accuracy

 

± 1

Physical Dimensions Body

H x W x D 14.8 x 8.1 x 6.4 cm

H x W x D 14.8 x 8.1 x 6.4 cm

Weight

900 g

900 g

Cable Length

1.9 m

1.9 m

Warranty

One Year

One Year

 

 