INPUT OPERATING MODES:
DIRECT, ANALOG TRIGGER, TTL TRIGGER, selectable with
toggle switch
FUNCTION MODES:
voltage source, current source, selectable with toggle
switch
SOURCE RANGES:
±1 mA [optional ±10 mA] (current source),
±100 V [optional ±45 V] (voltage source), selectable
with toggle switch
TIME RANGE FACTORS:
x10 µs, x100 µs, x1ms, selectable with toggle
switch, time range up to 99 ms
|
AMPLITUDE POTENTIOMETER:
3 digits, ±10 V, amplified internal with gain of 10
(voltage source), polarity selectable with toggle
switch
TIME POTENTIOMETER:
2 digits, 0 - 99 as multiplication factor selectable
INPUT VOLTAGE RANGE:
±10 V, amplified internally with gain of 10 (voltage
source) or gain of 1 (current source)
OUTPUT VOLTAGE:
±100 V max., isolated (optional ±45 V)
OUTPUT CURRENT:
±1 mA max., isolated (optional ±10 mA)
|