The LHBF-48X is
designed for processing small bio-electrical signals.
The input voltage range is ±12 V and the inputs are
protected up to ±100 V. Two overload-LEDs indicate
when the linear region of the amplifier is exceeded.
All inputs and the output are BNC-connectors.
The LHBF-48X is equipped with a differential input
stage with an AC/O/DC switch for each input and an
offset compensation. The gain stage allows the
amplification of the signal from 1-1000 in 1-2-5
steps. The gain is selected by an ten-position rotary
switch.
The amplified signal is applied to the output filter(s).
The eight-pole Bessel lowpass filter has sixteen
corner frequencies (30 Hz - 20 kHz). The output filter
can be bypassed by using a toggle switch. Optionally,
the LHBF-48X can be equipped with a single-pole Bessel
highpass filter with eleven corner frequencies (0.3 Hz
- 10kHz) and an OFF position. A NOTCH filter (50/60
Hz) is also available. The position of the GAIN and of
the FILTER switches can be read from monitor outputs
(1V/step). |