INPUT
RANGE:
±12 V; protected up to ±15 V
INPUT
IMPEDANCE:
10 MOhms related to ground
INPUT
CAPACITANCE:
30 pF
AC COUPLING AT
INPUT:
corner frequency of 0.1 Hz (if
AC/0/DC-switch at BNC connector is
switched to AC)
OFFSET
COMPENSATION:
±0.1 V, ±1 V or ±10 V, set by toggle
switch and ten-turn potentiometer
HIGHPASS
FILTER:
DPA 2FX
single pole, attenuation: -6 dB/octave,
corner frequencies (Hz): 3, 30, 100, 300,
400, 500, 600, 800, 1k, 3k, 10k
LOWPASS
FILTER:
DPA 2FX
single pole, attenuation: -6 dB/octave,
corner frequencies (Hz): 30, 100, 300,
500, 700, 900, 1.3k, 1.5k, 1.7k, 3k, 10k,
20k
|
HIGHPASS
FILTER:
DPA 2FS
single pole, attenuation: -6 dB/octave,
corner frequencies (Hz): OFF, 0.3, 5, 100,
300, 400, 500, 600, 800, 1k, 3k, 10k
LOWPASS
FILTER:
DPA 2FS
single pole, attenuation: -6 dB/octave,
corner frequencies (Hz): 30, 100, 300,
500, 700, 1k, 1.3k, 1.5k, 3k, 5k, 10k, 20k
GAIN:
rotary-switch: 1-2-5-10-20-50-100
GAIN RANGE:
selectable with toggle switch x1, x10,
x100
GAIN and GAIN RANGE are multiplied
giving a maximum amplification of 10000
OVERLOAD LEDS:
starting to light up at ±10 V
OUTPUT:
impedance 50 Ohms, max ±12 V, short
circuit protected
DIMENSIONS:
front panel 12 HP (60.6 mm) x 3U (128,5
mm), 7" (175 mm) deep. |